Agilent 7200 GC/Q-TOF Application Note: Analysis of Biomarkers in Crude Oil

This article describes the analysis of biomarkers in crude oil using the Agilent 7200 GC/Q-TOF system. With a high-resolution time-of-flight mass spectrometer, a diluted sample can be analyzed without fractionation, and the biomarkers of interest can be measured by exploiting the high selectivity of ion extraction at accurate mass. The system has excellent sensitivity, allowing for the selective detection of dibenzothiophene, alkylated dibenzothiophenes, and hopanes. Using the Agilent GC/Q-TOF system in MS/MS mode, low levels of steranes can be selectively detected.


Agilent 7200 GC/Q-TOF Application Note: Analysis of Biomarkers in Crude Oil - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
10
File format
PDF
File size
320 KB
Published
11 June, 2012
Updated
28 September, 2023
MD5 checksum
06E426CCD66279C702D05BC97B0473E0

About this manual

The document describes an analytical method for characterizing crude oil sources and identifying pollution sources by analyzing specific biological markers. The study utilizes a high-resolution time-of-flight mass spectrometer to analyze diluted samples directly without complex fractionation.

Key biomarkers analyzed include alkyl-dibenzothiophenes, hopanes (pentacyclic triterpanes), and steranes (sterol-derived polycyclic alkanes). The text details experimental conditions using an Agilent 7890A GC System combined with a 7200 Q-TOF system. It highlights the advantages of accurate mass determination and MS/MS mode for achieving high selectivity in complex matrices.

What's inside

  • Introduction
  • Experimental Conditions
  • Results: Dibenzothiophenes
  • Results: Hopanes
  • Results: Steranes
  • Conclusions

Frequently asked questions

What is the final concentration of the dibenzothiophene reference solution used for instrument performance checks?

The final concentration of the dibenzothiophene reference solution was 0.38 ng/pL.

What is the final oil concentration after diluting the crude oil extract in hexane?

The final oil concentration after dilution was 1 mg/mL.

What are the dimensions of the DB-5MS column used in the GC system?

The DB-5MS column used was 30 m x 0.25 mm with a 0.25 pm film thickness.