Agilent 5060-9084 Moisture Trap User Guide

This product is a moisture trap, which is mainly used to remove the trace amount of water present in nearly all carrier gases.


Agilent 5060-9084 Moisture Trap User Guide - cover page
Brand
Agilent
Document type
User Manual
Language
English
Pages
4
File format
PDF
File size
120 KB
Published
08 June, 2012
Updated
30 September, 2023
MD5 checksum
9B2B5A1D4A336E4EE3738387ED1C5436

About this manual

The document describes the Agilent 5060-9084 Moisture Trap, which is packed with Molecular Sieve 5A (45/60 mesh) and fitted with 1/8-inch OD Swagelok fittings. It is designed to remove trace amounts of water from carrier gases used in Agilent Gas Chromatographs. The trap can be installed directly between the gas line regulator and the instrument inlet.

What's inside

  • General Information
  • Conditioning
  • Installation
  • Regeneration
  • Repacking

Specifications

Packing MaterialMolecular Sieve 5A, 45/60 mesh
Fitting Type1/8-inch OD Swagelok fittings
Flow Pressure Dropapproximately 2 psig

Frequently asked questions

How do you condition the trap?

Remove caps, attach to gas supply, set flow rate at 60 mL/min and oven temperature at 350°C, then condition overnight.

When should the trap be regenerated?

Regeneration should be done periodically after using one to four cylinders of gas depending on grade.