Varian 971-FP Flash Purification Software Upgrade Instructions

This document describes how to upgrade the Varian IntelliFlash software and firmware on the 971-FP and IF310 instruments. These versions of software and firmware address known bugs and offer new features.


Varian 971-FP Flash Purification Software Upgrade Instructions - cover page
Brand
Agilent
Document type
Service Manual
Language
English
Pages
10
File format
PDF
File size
643 KB
Published
07 June, 2012
Updated
29 September, 2023
Model
971-FP
MD5 checksum
6BF46DE0CD9BE719AE4D84956A4B36AB

About this manual

This document provides step-by-step instructions for performing a software upgrade on the Varian 971-FP Flash Purification system. The process involves uninstalling existing software, transferring new installation files to a USB drive, and running the setup wizard on the instrument.

The guide also details how to update firmware for connected modules such as the Fraction Collector and UV detector. Specific procedures are included for entering bootloader mode on the UV module if automatic updates fail, as well as instructions for re-installing drivers for optional ELSD (Evaporative Light Scattering Detector) USB-Serial cables.

What's inside

  • Introduction
  • Software Installation
  • Firmware Update
  • UV Detector Bootloader Mode
  • ELSD Driver Installation

Frequently asked questions

Which instruments are supported by this software upgrade?

This document describes how to upgrade the Varian IntelliFlash software and firmware on the 971-FP and IF310 instruments.

Can the IntelliFlash software be run on external computers?

No, running the IntelliFlash software on external laptop or desktop computers is not supported.

What happens to historic data files after installation?

All historic data files will be lost following software installation unless they are exported beforehand.