This document describes how to upgrade the Varian IntelliFlash software and firmware on the 971-FP and IF310 instruments. These versions of software and firmware address known bugs and offer new features.
This document provides step-by-step instructions for performing a software upgrade on the Varian 971-FP Flash Purification system. The process involves uninstalling existing software, transferring new installation files to a USB drive, and running the setup wizard on the instrument.
The guide also details how to update firmware for connected modules such as the Fraction Collector and UV detector. Specific procedures are included for entering bootloader mode on the UV module if automatic updates fail, as well as instructions for re-installing drivers for optional ELSD (Evaporative Light Scattering Detector) USB-Serial cables.
This document describes how to upgrade the Varian IntelliFlash software and firmware on the 971-FP and IF310 instruments.
No, running the IntelliFlash software on external laptop or desktop computers is not supported.
All historic data files will be lost following software installation unless they are exported beforehand.