Varian VS C15 Leak Detector Filament Replacement Field Instructions

This document is about VS C15 leak detector filament replacement instruction, which introduces the operation steps of replacing filaments


Varian VS C15 Leak Detector Filament Replacement Field Instructions - cover page
Brand
Agilent
Document type
Service Manual
Language
English
Pages
15
File format
PDF
File size
804 KB
Published
07 June, 2012
Updated
30 September, 2023
Model
C15
MD5 checksum
BC30CC11CDDD621DC52AAAD7B27DCEEF

About this manual

This document provides field installation instructions for performing filament replacement on a Varian VS C15 Leak Detector. It includes important notices regarding the manufacturer's transition from Varian to Agilent Technologies and defines standard documentation symbols such as NOTE, CAUTION, and WARNING.

The manual outlines specific safety precautions, including handling solvents, O-ring care, and static electricity protection. It details the required equipment, such as metric Allen wrenches and torque wrenches, and provides a step-by-step installation procedure for accessing and replacing the internal components.

What's inside

  • Preface
  • Hazard and Safety Information
  • Solvents
  • O-ring Care
  • Equipment Required
  • Installation Procedure

Frequently asked questions

What tools are required to open the unit's top cover?

You need an M4 Allen wrench and a Philips head screwdriver.

How long must you wait after disconnecting power before opening the unit?

You must wait 30 seconds for the high voltage to dissipate.

What specific grease is recommended for O-rings that contact vacuum surfaces?

Apply a small amount of Apiezon L grease and wipe it shiny dry.

Download PDF (804 KB)

More Agilent Manuals


View all Agilent manuals →