Agilent Technologies 979 Detector NW25 Calibrated Leak: Instructions

This document introduces the use of calibrated helium leak and provides detailed step-by-step instructions


Agilent Technologies 979 Detector NW25 Calibrated Leak: Instructions - cover page
Brand
Agilent
Document type
User Manual
Language
English
Pages
6
File format
PDF
File size
177 KB
Published
07 June, 2012
Updated
01 October, 2023
Model
NW25
MD5 checksum
A94FAFCBF8A5215EC25AFB8E286AB246

About this manual

This document provides instructions and technical information for the Varian NW25 Calibrated Helium Leak. The device is designed with an integral helium reservoir and an isolating valve, featuring either an NW25 or a 1-1/8 inch diameter sealing surface to fit the test port of any Varian Leak Detector.

The manual explains the principles of operation, noting that helium permeates through the pyrex-glass finger. It details how leak rates are affected by temperature and depletion over time, providing a table of typical depletion rates for various PIN suffixes. Users are advised to leave the valve open when not in use to maintain stability.

What's inside

  • Introduction
  • Use
  • Principles of Operation
  • Depletion Rates
  • Temperature Effects
  • Technical Assistance

Frequently asked questions

What sealing surfaces does the NW25 calibrated leak fit?

It is furnished with either an NW25 or a 1-1/8 inch diameter sealing surface to fit the test port of any Varian Leak Detector.

How should the calibrated leak be stored when not in use?

When the calibrated leak is not in use, it should always be left with the valve open and stored at the same temperature at which it will be used.

How much does the leak rate increase per degree Celsius?

The actual leak rate increases with increased temperature at 3% per degree (Celcius).