This document is the Olympus AU 400 - 600 - 640 user manual.
This document provides detailed guidelines for operating the Olympus AU 400/600/640 automatic clinical analyzer. It describes the recommended procedures for reagent handling, sample preparation, and instrument settings to ensure accurate test results.
The manual lists applications for various assays (e.g., acid phosphatase, albumin, alkaline phosphatase, amylase, uric acid, zinc), specifies preparation steps, storage temperatures, stability periods, measurement wavelengths, and calibration factors. It also includes information on dilution rates, linearity, and on‑board stability periods for each test.
| On‑board stability period | 30 days |
|---|---|
| Reagent storage temperature | 2‑8 °C |
| Dilution pre‑dilution rate | 1 |
Reagents should be refrigerated at 2‑8 °C.
The pre‑dilution rate is 1 for all listed assays.
The primary wavelength for the ALP test is 410 nm.