This document is the instruction manual for Philips heart defibrillator M5528A and M5529A. It mainly introduces the risk of smoke or fire caused by incorrect wiring during user installation, and proposes preventive measures.
The M5528A HeartStart MRx Vehicle Wall Mount is designed to hold the HeartStart MRx Monitor/Defibrillator (M3536A) during transport, typically within an ambulance. Installation should be performed by a qualified mechanical and electrical technician, and the mount must be screwed to the vehicle chassis.
The guide covers preparing the monitor, inserting and removing it from the mount, performance verification, safety warnings, maintenance, repair, and product specifications. It also includes a field safety notice warning that unauthorized modifications to the wall mount, cables, mounting brackets, or DC Power Module can create a smoke or fire hazard.
| Standards | IEC 60601-1 (2001), IEC 60601-1-2 (2001), EN 1789 (1999) |
|---|---|
| Weight | 1.5 kg (3.3 lb), excluding cables |
| Dimensions | 75 x 300 x 320 mm (3.0 x 11.2 x 12.6 in) |
| Operating temperature | 0 to 40 C (32 to 104 F) |
| Storage temperature | -30 to 70 C (-22 to 158 F) |
| Humidity | 5% to 95%, non-condensing |
| Altitude | -500 to 7500 m (-1,500 to 25,000 ft) |
The M5528A HeartStart MRx Vehicle Wall Mount is designed to hold the HeartStart MRx Monitor/Defibrillator (M3536A) during transport.
No, the HeartStart MRx cannot be inserted into the Vehicle Wall Mount when the rear accessory pouch, bedrail hook, or M3539A AC power module is attached.
The installation should be verified every month.
Stop using the Vehicle Wall Mount and correct the installation.