RayBiotech MIP-1 beta/CCL4 IQELISA Kit User Manual

RayBiotech MIP-1 beta/CCL4 IQELISA Kit User Manual

User Manual 11 pages · 155 KB · EN

This document provides detailed instructions for using the RayBio® Mouse MIP-1 beta/CCL4 IQELISA Kit, which combines ELISA specificity with qPCR sensitivity. It includes reagent preparation, assay procedure steps, and data calculation methods using Ct values. The kit is designed for use with real-time PCR instruments and requires specific materials and storage conditions for optimal performance.

The manual covers storage guidelines, reagent dilutions, standard preparation, and troubleshooting. It also explains how to calculate results using log-log graphs or linear regression, with formulas for determining unknown sample concentrations based on calibration curves.

Read online

RayBiotech CCL4 IQELISA Kit specifications

Target AnalyteHuman MIP-1 beta
Sample Typesserum, plasma, and cell culture supernatants
Plate Type96-well PCR plate
Minimum Quantifiable Dose0.97 pg/ml
Serum Spike Recovery86% (range 80% to 91%)
Intraplate CVbelow 10%
Interplate CVbelow 15%
Storage Temperature2° to 8°C

Source: RayBiotech Mouse MIP-1 beta/CCL4 IQELISA Kit User Manual

All RayBiotech CCL4 IQELISA Kit documents by type

User Manual 3

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

What is the minimum quantifiable dose of MIP-1 beta in this kit?

The minimum quantifiable dose is typically 0.97 pg/ml.

From: RayBiotech MIP-1 beta/CCL4 IQELISA Kit User Manual

What samples can be used to measure Human MIP-1 beta?

The assay is designed for the quantitative measurement of Human MIP-1 beta in serum, plasma, and cell culture supernatants.

From: RayBiotech MIP-1 beta/CCL4 IQELISA Kit User Manual

What is the minimum quantifiable dose of MIP-1 beta?

The minimum quantifiable dose is typically 0.97 pg/ml.

From: RayBiotech Mouse MIP-1 beta/CCL4 IQELISA Kit User Manual

What sample types can be tested with this kit?

The kit is designed for serum, plasma, and cell culture supernatants.

From: RayBiotech Mouse MIP-1 beta/CCL4 IQELISA Kit User Manual

What is the minimum quantifiable dose of MIP-1 beta in this kit?

The minimum quantifiable dose is typically 0.97 pg/ml.

From: RayBiotech Mouse MIP-1 beta/CCL4 IQELISA Kit User Manual

What is the recovery rate for serum spike tests?

The recovery rate is 86%, with a range from 80% to 91%.

From: RayBiotech Mouse MIP-1 beta/CCL4 IQELISA Kit User Manual