Fluke 6331 User Guide Supplement

Fluke 6331 manual supplement (Issue 4) with changes to environmental conditions, concrete floor installation, cutout reset and maintenance instructions.


Fluke 6331 User Guide Supplement - cover page
Brand
Fluke
Category
Calibration Components
Document type
Other
Language
English
Pages
3
File format
PDF
File size
182 KB
Published
18 November, 2026
Updated
18 November, 2026
Model
6331
MD5 checksum
2F0EAD65BB606F3B141F2D3C3F811F7B

About this manual

This is a manual supplement for the Fluke 6331, issued to keep the 6331 Users Guide accurate. It is dated 3/21 and carries Issue 4 of the supplement; the original guide was printed in January 2013 and distributed as an electronic manual on CD-ROM (CD PN: 3750900). The supplement lists numbered changes that add to, replace or delete passages on specific pages of the guide.

The changes cover a note on control sensitivity to electromagnetic and conducted interference in the Environmental Conditions section, a new procedure for installing the unit on a concrete floor with a bracket and anchor bolts, a deletion in the Unpacking section, a rewritten Reset Cutout and Cutout description, removal of the cutout reset mode parameter and its menu entries, a replacement for section 9.16.1 CTO, a replacement maintenance bullet on checking the over-temperature cutout, and a replacement troubleshooting bullet for a flashing "Cut-out" display.

What's inside

  • Environmental Conditions
  • Installation On A Concrete Floor
  • Unpacking
  • Reset Cutout
  • Cutout
  • Operating Parameters
  • CTO
  • Operating Parameters Menu
  • Maintenance
  • Causes and Solutions

Specifications

Radiated EM disturbance frequency range250 MHz to 400 MHz
Stability change for radiated disturbance above 1 V/m to 3 V/mAdd 0.0025 °C to the stability specification
Conducted disturbance frequency range8 MHz to 80 MHz
Stability change for conducted disturbance above 3 VAdd 0.005 °C to the stability specification
Floor hole size for bracket installation1/4 inch x 1 3/4 inch deep (approximately 6.5 mm x 32 mm)
Air circulation clearanceMinimum of 6 inches
Cutout set-point adjustment10 °C or 20 °C above the maximum bath operating temperature
Over-temperature cutout check intervalEvery 6 months

Frequently asked questions

How often should the over-temperature cutout be checked?

It should be checked every 6 months to see that it is working properly.

What happens if the cutout is activated by excessive bath temperature?

Power to the heater is shut off, the bath cools, and the display flashes Cut-out until the temperature falls below the reset temperature and the cutout is reset.

How is the over-temperature cutout calibrated (CTO)?

It is not adjustable by software; it is adjusted with an internal potentiometer and is set at the factory.

What clearance is needed when installing on a concrete floor?

The bracket must be installed so the bath has a minimum of 6 inches of clearance for air circulation.