Fluke 6331 manual supplement (Issue 4) with changes to environmental conditions, concrete floor installation, cutout reset and maintenance instructions.
This is a manual supplement for the Fluke 6331, issued to keep the 6331 Users Guide accurate. It is dated 3/21 and carries Issue 4 of the supplement; the original guide was printed in January 2013 and distributed as an electronic manual on CD-ROM (CD PN: 3750900). The supplement lists numbered changes that add to, replace or delete passages on specific pages of the guide.
The changes cover a note on control sensitivity to electromagnetic and conducted interference in the Environmental Conditions section, a new procedure for installing the unit on a concrete floor with a bracket and anchor bolts, a deletion in the Unpacking section, a rewritten Reset Cutout and Cutout description, removal of the cutout reset mode parameter and its menu entries, a replacement for section 9.16.1 CTO, a replacement maintenance bullet on checking the over-temperature cutout, and a replacement troubleshooting bullet for a flashing "Cut-out" display.
| Radiated EM disturbance frequency range | 250 MHz to 400 MHz |
|---|---|
| Stability change for radiated disturbance above 1 V/m to 3 V/m | Add 0.0025 °C to the stability specification |
| Conducted disturbance frequency range | 8 MHz to 80 MHz |
| Stability change for conducted disturbance above 3 V | Add 0.005 °C to the stability specification |
| Floor hole size for bracket installation | 1/4 inch x 1 3/4 inch deep (approximately 6.5 mm x 32 mm) |
| Air circulation clearance | Minimum of 6 inches |
| Cutout set-point adjustment | 10 °C or 20 °C above the maximum bath operating temperature |
| Over-temperature cutout check interval | Every 6 months |
It should be checked every 6 months to see that it is working properly.
Power to the heater is shut off, the bath cools, and the display flashes Cut-out until the temperature falls below the reset temperature and the cutout is reset.
It is not adjustable by software; it is adjusted with an internal potentiometer and is set at the factory.
The bracket must be installed so the bath has a minimum of 6 inches of clearance for air circulation.