Thermo Fisher Scientific Summit spectrometer site and safety guide covering power, temperature, humidity, purge gas, vibration, and laser safety.
The Thermo Fisher Scientific Summit Site and Safety Guide provides pre-installation and safety information for the Summit spectrometer. It describes how to inspect and move the instrument, workspace requirements, electrical service specifications, and acceptable temperature, humidity, storage, vibration, and magnetic-field conditions.
The guide also covers purge requirements and gas selection, hazardous and corrosive materials, fire and burn precautions, biological and radioactive material handling, laser safety, and cleaning. It states that the Summit weighs 9.6 kg and has dimensions of 33.8 cm wide, 24.0 cm high, and 32.3 cm deep. This manual is written in Chinese.
| Weight | 9.6 kg (21 lb) |
|---|---|
| Dimensions | 33.8 cm W × 24.0 cm H × 32.3 cm D |
| Input current | 1.5 A max |
| Input voltage | 100 to 240 VAC |
| Line frequency | 50 to 60 Hz |
| Operating temperature | 15°C to 35°C |
| Operating humidity | 20% to 80%, non-condensing |
| Storage temperature | -20°C to 60°C |
| Maximum storage humidity | 85% RH, non-condensing |
| Operating altitude | Up to 2,000 m (6,500 ft), indoor use |
| Laser wavelength | 850 nm |
| Laser maximum power | 0.39 mW |
| Laser class | Class 1 (FDA-CDRH and IEC60825-1:2014) |
The Summit requires 100 to 240 VAC.
Normal operation is supported from 15°C to 35°C.
Use only dry air or nitrogen.
The laser source is an 850 nm diode.