Installation instructions for the Thermo Fisher Scientific UV-Retrofit-Kit used on MSC Advantage MSC 1.2 and MSC 1.8 safety cabinets.
These installation instructions cover fitting the UV-Retrofit-Kit to the Thermo Fisher Scientific MSC Advantage safety cabinet, models MSC 1.2 and MSC 1.8. They cover general safety notes, warranty conditions, symbols, installation safety requirements, applicable standards, required tools and fasteners, and the installation sequence from removing covers through testing the UV lamp.
The kit includes a UVC lamp, ballast unit, lamp sockets, wiring diagrams and wire sets for the MSC 1.2 and MSC 1.8 light boxes, along with mounting hardware and a nameplate. The document also lists the kit components and technical data for the MSC 1.2 and MSC 1.8.
| Applicable safety cabinet models | MSC 1.2 and MSC 1.8 |
|---|---|
| UVC lamp power | 36 W |
| UVC lamp length | 826 mm |
| Ballast unit rating | 2 X 32 W 120 - 277 V |
| Lamp socket type | Rotating socket for UVC lamp, G5 |
| Mains connection | AC, 230 V - 50-60 Hz |
| Protection class | I |
| Protection type | IP 20 |
| Cable length | 3.7 m |
| Circuit breaker | B16 / Fuse T 16 A |
| Standards and directives | NF 095 Rev3.2006, IEC 61010-1:2002-08, EN 61010-1:2002, EN 12469:2000, DIN EN 61326-1:2008-06, Low Voltage Directive 2006/95/EC, EMV Directive 2004/108/EC |
The instructions apply to the MSC Advantage safety cabinet models MSC 1.2 and MSC 1.8.
Repairs must be carried out only by appropriately qualified personnel, and electrical work only by electrical expert personnel. Defined spare part replacement must be performed by Thermo Fisher Scientific Technical Service or authorized, trained service personnel.
The external gas supply must be shut off prior to installation, the work room must be sufficiently ventilated, and the exhaust system should be activated if equipped.
Any removed or disabled safety devices must be reinstalled and checked for proper operation, and a test run must be performed before startup and operation.