Kistler 4049B...DS Water Cooled Absolute Pressure Sensor User Manual

User manual for Kistler's water-cooled absolute pressure sensor, designed for high-temperature gas pressure measurement in exhaust systems.


Kistler 4049B...DS Water Cooled Absolute Pressure Sensor User Manual - cover page
Brand
Kistler
Document type
User Manual
Language
English
Pages
4
File format
PDF
File size
654 KB
Published
02 November, 2026
Updated
02 November, 2026
MD5 checksum
18FDC4E31554ADA60E973018A627F80D

About this manual

This document provides installation, maintenance, and technical specifications for the Kistler Type 4049B...DS water-cooled absolute pressure sensor. It is designed for high-temperature gas pressure measurement, especially in exhaust systems of internal combustion engines. The sensor features a rugged design, digital temperature compensation, and an integrated water cooling system to manage thermal effects. Installation requires specific tools and torque settings, and maintenance may involve cleaning or replacing the protector screen to prevent soot buildup.

Technical data includes measuring ranges (0–5 bar and 0–10 bar), electrical connection type (Fischer connector), amplifier compatibility (Type 4665, 4665B, 4624A), and cooling requirements (0.3–0.5 l/min water flow). The sensor is rated for temperatures up to 1100°C and has a protective class of IP65.

What's inside

  • Description
  • Application
  • Installation
  • Maintenance
  • Technical Data
  • Figures and Drawings

Specifications

Measuring range0 ... 5 bar / 0 ... 10 bar
Overload15 bar / 25 bar
Electrical connectionFischer connector 5 pole (S103A054)
Amplifier compatibilityType 4665, 4665B, 4624A
Reference temperature60 °C
Sensor temperature range0/120 °C
Temperature compensationDigital
Max. deviation pressure≤±0.3 % FSO
Max. deviation temperature≤±3 °C
Natural frequency>60 kHz
Required cooling flow0.3 ... 0.5 l/min
Sensor mounting torque20 N·m
Screen mounting torque3 N·m
Weight (without connector)30 g
Protective classIP65

Frequently asked questions

What is the maximum operating temperature of the sensor?

The sensor can operate in temperatures exceeding 1100°C due to its integrated water cooling system.

What tools are required for installation?

The sensor requires specific tools including mounting tool Type 1300A19 and torque wrench Type 1300A39, with a tightening torque of 20 N·m.

What is the sensor's measuring range?

The sensor is available in two measuring ranges: 0–5 bar and 0–10 bar.

What is the sensor's protective class?

The sensor has an IP65 protective class.

Download PDF (654 KB)

More Kistler Manuals


View all Kistler manuals →