BD CBA Human Active Caspase-3 kit (Cat. No. 552124) instructions for measuring active caspase-3 protein in cell lysates by flow cytometry.
This manual describes the BD Cytometric Bead Array (CBA) Human Active Caspase-3 kit, Cat. No. 552124. The kit uses a single bead population coated with a capture antibody specific for caspase-3, a PE-conjugated detection antibody, and a lyophilized cell lysate standard to form sandwich complexes that are read on a flow cytometer and analyzed with BD CBA Software. The kit provides reagents for the quantitative analysis of 100 samples, including up to three standard curve sets, and is intended for research use only.
The document covers the principle of the test, reagents provided and materials required but not provided, preparation of the lysate standard, cell lysis buffer mixture and test samples, the assay procedure, cytometer setup with BD FACSComp software and Cytometer Setup Beads, data acquisition and analysis, typical data, performance (sensitivity, specificity, precision), and troubleshooting tips. Storage instructions and hazardous ingredient information for the kit components are also given.
| Storage temperature of kit components | 4 °C |
|---|---|
| Human Caspase-3 Capture Beads | 1 vial, 5 ml (100-test vial) |
| Human Active Caspase-3 PE Detection Reagent | 1 vial, 5 ml; 50 µl/test |
| Human Active Caspase-3 Lysate Standard | 3 vials, 0.2 ml lyophilized; 30,000 Units/ml after reconstitution |
| Cytometer Setup Beads | 1 vial, 1.5 ml; 30 tests, sufficient for 10 setup procedures |
| Wash Buffer | 1 bottle, 130 ml (PBS solution, 1×) |
| Assay Diluent | 1 bottle, 30 ml (1×) |
| Number of samples supported | 100 samples |
| Required flow cytometer | 488 nm laser, detecting emissions at 576 and 670 nm |
| Protease Inhibitor Cocktail | 1 vial, 1 ml lyophilized; reconstituted to 50× bulk |
It is used to quantitatively measure active caspase-3 protein levels in a single cell lysate sample, using flow cytometry.
The kit provides sufficient reagents for quantitative analysis of 100 samples, including generation of up to three standard curve sets.
A flow cytometer with a 488 nm laser that can detect fluorescence emissions at 576 and 670 nm, such as a BD FACScan or BD FACSCalibur instrument.
Store them at 4 °C and do not freeze; vortex the vial vigorously for 3–5 seconds before taking an aliquot because the beads settle out of suspension.
Each vial is reconstituted in 0.2 ml of deionized water to prepare a 5× bulk standard, which should be stored at 2–8 °C and used within 12 hours.