BD FACS Canto II Flow Cytometer Quick User Guide

Describes how to start and shut down the BD FACS Canto II Flow Cytometer


BD FACS Canto II Flow Cytometer Quick User Guide - cover page
Brand
B&D
Document type
Quick Start Guide
Language
English
Pages
8
File format
PDF
File size
540 KB
Published
20 October, 2023
Updated
20 October, 2023
MD5 checksum
02C1DA5F85D2B390539BA91E02ECDFAA

About this manual

This guide provides step-by-step instructions for starting up and shutting down the BD FACS Canto II flow cytometer. It includes procedures for connecting the instrument to the computer, initializing fluidics, and logging in. It also covers troubleshooting common issues such as laser power warnings and empty fluid containers.

The guide emphasizes safety and proper maintenance, including replacing waste and sheath fluid containers, priming the system after refills, and ensuring the machine is properly shut down to avoid connectivity issues for the next user.

What's inside

  • FACS Canto II Start Up Procedure
  • FACS Canto II Shut Down Procedure
  • Infrequently Encountered Problems

Frequently asked questions

What should I do if the machine does not connect or shows persistent laser low power warnings?

Reboot the system: exit FACSDiva software, shut down the computer, turn off the instrument, restart the instrument and reboot the computer, then repeat the start-up procedure.

How long does the Fluidics Startup take?

The Fluidics Startup takes 5-7 minutes.

How long should you wait for the lasers to reach power?

It is recommended that you allow 30 minutes for the lasers to get up to power.

How heavy are the full Cubitainers?

Full Cubitainers weigh approximately 20 kg.