Ohaus Scout SE Series Electronic Balance Operation Manual

The SE series electronic balance (Scout™) is an electronic balance with installation, operation and maintenance functions. It is used in a dry and stable environment and can perform operations such as zeroing/taring, printing, calibration, etc. It also supports multiple weighing units and counting mode.


Ohaus Scout SE Series Electronic Balance Operation Manual - cover page
Brand
OHAUS
Category
Other Electronics
Document type
User Manual
Language
Chinese
Pages
16
File format
PDF
File size
288 KB
Updated
29 September, 2023
MD5 checksum
F00C52C3BCF30550263A19A280FC4AB8

About this manual

This manual covers the Scout SE series electronic balances, describing safety precautions, unpacking, installation, power options, and calibration procedures. It explains the operation panel functions, menu structure, and how to select weighing units and modes.

Technical specifications for each model are listed, along with optional accessories such as RS‑232 and USB communication modules, density measurement kits, and calibration weights. Contact information for service and warranty is provided at the end of the document. This manual is written in Chinese.

What's inside

  • Introduction
  • Safety
  • Installation
  • Operation Panel
  • Range Calibration
  • Weighing Modes
  • Menu Structure
  • Technical Parameters
  • Accessories

Specifications

Maximum weighing capacity (g)200–1500
Minimum increment (g)0.01
Accuracy (MPE) for 0–50 g±0.05 g
Range calibration weight (g)200
Operating temperature range10–30 °C

Frequently asked questions

How do I perform a range calibration?

Press the short‑press clear/tare key to power on, then long‑press the print/calibrate key; when the display shows [CAL] release the key, place the required calibration weight, and short‑press Yes to complete the calibration.

What power sources can be used with the balance?

The balance can be powered from the supplied AC adapter or from four AA batteries (batteries are not included).

How can I change the weighing unit?

Short‑press the unit conversion / weighing mode button to cycle through the available units such as g, kg, etc.

How is the automatic shutdown feature configured?

In the Setup menu, set the Auto‑Off option to On; the balance will shut down after four minutes of inactivity.