CP series electronic balance instruction manual, introduces the functions, installation, use and maintenance of the balance
This manual provides step‑by‑step instructions for safely installing and operating Ohaus CP series electronic balances. It includes safety precautions, unpacking, leveling, power connection, initial calibration and detailed weighing modes such as count, percent and pendant weighing.
The guide also describes menu navigation, calibration procedures, unit customization, RS‑232 communication, and lists technical specifications for each model in the CP series, along with maintenance and service information. This manual is written in Chinese.
| Maximum weighing capacity | 65 g – 4100 g (depending on model) |
|---|---|
| Actual graduation (resolution) | 0.0001 g |
| Power supply voltage | 8‑14.5 VAC, 50/60 Hz (or 8‑20 VDC, 4 W) |
| RS‑232 interface | Available on most models |
| External dimensions (W×D×H) | 196 mm × 320 mm × 310 mm |
Press the Unit key to open the .Unit. menu, select SET ON, then use the Confirm and Cancel keys to adjust the coefficient, multiplier and least significant digit for the custom unit.
Pin 2 = TxD, Pin 3 = RxD, Pin 5 = GND, Pin 7 = CTS, Pin 8 = RTS.
In the Settings menu, enable the Auto‑Zero option (AzsMM) by setting it to ON; the balance will automatically zero when environmental conditions change.