Portable gas leak detector, model C16, for home and office use.
This O&M manual covers the PortaSens II Model C16, a hand-held gas leak detector from Analytical Technology, Inc. (ATI) used to detect leaks of toxic gas and oxygen displacement or deprivation. The device adapts to different gases by inserting interchangeable H10 Smart Sensor modules, which retain calibration and configuration settings for the target gas. It uses a battery-powered intake pump with a flexible sampling wand, and displays readings on a backlighted graphics LCD with four soft keys and an alarm beeper.
The manual describes package contents and reference drawings, operation and alarm behavior, device configuration, zero and span calibration, and the data logger. It also documents the C16Log Windows application for transferring logged readings to a PC as ASCII csv files, along with maintenance of the intake pump filter and a troubleshooting table.
| Gas Sensor | Accepts H10 Smart Sensor gas sensor modules (not included) |
|---|---|
| Display | Backlighted, graphics LCD |
| Accuracy | Sensor dependent, typically ±5% of value |
| Sensitivity | 1% of sensor module range |
| Repeatability | ±1% of sensor module range |
| Input/Output | RS-232 for logged readings; 0-1 VDC analog (optional cable) |
| Data Logger | 1, 5, 10, and 15 minute intervals; 12,000 samples |
| Alarms | Three level alarms, low battery alarm, low flow rate alarm |
| Power | D-cell alkaline battery, 75 hours; internal rechargeable backup, 6 hours |
| Operating Temperature | -25º to +55º C |
| Humidity | 0-95% non-condensing |
| Construction | Glass filled nylon, PVC, and stainless steel |
| Shipping Weight | 3 lbs. (1.4 kg) |
The primary D-cell alkaline battery provides about 75 hours of operation, and the internal rechargeable backup battery provides about 6 hours.
It accepts H10 Smart Sensor gas sensor modules, which are ordered separately and retain calibration and configuration settings for the target gas.
The data logger stores 12,000 samples, which is 8 days at the 1-minute interval, with 1, 5, 10, and 15 minute intervals available.
Press and hold all four front panel keys for 5 seconds to reset and turn off the device, then start it again with the green button.