This document is an operation manual that introduces the features, specifications, installation and usage of a self-cleaning membrane dissolved oxygen system
This manual provides operation and maintenance instructions for the ATI Q46D Auto-Clean Membrane Dissolved Oxygen System. The system is designed for continuous online monitoring of dissolved oxygen in solutions, utilizing a galvanic membrane sensor and an integrated air-blast cleaning assembly known as the Q-Blast. It is suitable for water streams with temperatures ranging from 0 to 50 °C and offers a full-scale operating range of 0 to 40 ppm.
| Main Parameter Range | 0.00 to 40.00 ppm or 0.00 to 40.00 mg/L |
|---|---|
| Power Supply | 90-260 VAC or 12-24 VDC |
| Analog Outputs | Two 4-20 mA outputs |
| Relay Outputs | Three SPDT relays |
| Accuracy | ±0.05 PPM or 0.5% of Span |
| Response Time | 180 seconds to 90% of step input |
| Temperature Input | Pt1000 RTD |
The default cleaning frequency is once every 24 hours, though this can be increased if needed for specific applications.
The maximum sensor cable length is 1000 feet (300 m) when using a junction box.
The analyzer service operates in ambient temperatures from -20 to 60 °C (-4 to 140 ºF).