This document describes the technical parameters, operating instructions and maintenance instructions for the ALCATEL 22004A-2 vacuum pump.
This manual covers the Alcatel 2004 A mechanical (primary) vacuum pump, a two-vane, two-stage rotary pump available with three-phase motors (220/380 V, 50 Hz, or 220/440 V, 60 Hz) or single-phase motors (220 V, 50 Hz, or 115/230 V, 60 Hz). It is written in French, German and English.
The manual describes first starting and normal starting, inlet and outlet connections (NW 25 Pneurop flanges), the manual gas ballast valve, oil filling with Alcatel VP 1 oil, and maintenance procedures including oil draining, flushing, dismantling, re-assembly and ordering spare parts. An addenda section covers the mist eliminator and automatic gas ballast accessories.
| Pump type | Vaned pump, two stages |
|---|---|
| Nominal rotational speed | 1500 rpm (50 Hz) / 1800 rpm (60 Hz) |
| Free air displacement | 4.5 m3/h (50 Hz) / 5.4 m3/h, 3.1 CFM (60 Hz) |
| Air displacement (Pneurop method) | 4 m3/h (50 Hz) / 4.8 m3/h, 2.8 CFM (60 Hz) |
| Base pressure (without gas ballast) | < 1.10-4 mbar |
| Oil capacity | 0.75 liter |
| Motor power | 0.37 kW (1/2 HP) |
Alcatel recommends VP 1 oil, with which the published characteristics are measured. Shell Talpa 30, Antar PV 600 and Coditec PV 195 A oils may also be used; for tough applications a synthetic oil such as ELF-BARELF 49 is recommended.
Yes. Since the pump is airtight, no precautions are needed for cold starts provided the temperature is not below 10 °C (50 °F).
Yes, the pump can be stopped without introducing air because it is airtight and oil migration is controlled by an anti-suckback valve.
Momentarily apply power and check that the end of the shaft turns in the direction of the arrow on the frame; the pumps are delivered wired for the maximum voltage.