ALCATEL 22004A-2 Vacuum Pump User Manual

This document describes the technical parameters, operating instructions and maintenance instructions for the ALCATEL 22004A-2 vacuum pump.


ALCATEL 22004A-2 Vacuum Pump User Manual - cover page
Brand
Alcatel
Category
Other products
Document type
User Manual
Language
English
Pages
26
File format
PDF
File size
5.5 MB
Published
08 November, 2023
Updated
08 November, 2023
MD5 checksum
6563BAF6ECDAB9D3D55E101A06317FCB

About this manual

This manual covers the Alcatel 2004 A mechanical (primary) vacuum pump, a two-vane, two-stage rotary pump available with three-phase motors (220/380 V, 50 Hz, or 220/440 V, 60 Hz) or single-phase motors (220 V, 50 Hz, or 115/230 V, 60 Hz). It is written in French, German and English.

The manual describes first starting and normal starting, inlet and outlet connections (NW 25 Pneurop flanges), the manual gas ballast valve, oil filling with Alcatel VP 1 oil, and maintenance procedures including oil draining, flushing, dismantling, re-assembly and ordering spare parts. An addenda section covers the mist eliminator and automatic gas ballast accessories.

What's inside

  • Technical characteristics
  • Description
  • First starting
  • Normal starting
  • Inlet connections
  • Manual gas ballast valve
  • Maintenance
  • Draining of the oil
  • Dismantling and re-assembly
  • Ordering spare parts

Specifications

Pump typeVaned pump, two stages
Nominal rotational speed1500 rpm (50 Hz) / 1800 rpm (60 Hz)
Free air displacement4.5 m3/h (50 Hz) / 5.4 m3/h, 3.1 CFM (60 Hz)
Air displacement (Pneurop method)4 m3/h (50 Hz) / 4.8 m3/h, 2.8 CFM (60 Hz)
Base pressure (without gas ballast)< 1.10-4 mbar
Oil capacity0.75 liter
Motor power0.37 kW (1/2 HP)

Frequently asked questions

What oil should be used in the Alcatel 2004 A pump?

Alcatel recommends VP 1 oil, with which the published characteristics are measured. Shell Talpa 30, Antar PV 600 and Coditec PV 195 A oils may also be used; for tough applications a synthetic oil such as ELF-BARELF 49 is recommended.

Can the pump be started cold?

Yes. Since the pump is airtight, no precautions are needed for cold starts provided the temperature is not below 10 °C (50 °F).

Can the pump be stopped without introducing air?

Yes, the pump can be stopped without introducing air because it is airtight and oil migration is controlled by an anti-suckback valve.

How is the direction of rotation checked on three-phase models?

Momentarily apply power and check that the end of the shaft turns in the direction of the arrow on the frame; the pumps are delivered wired for the maximum voltage.

Download PDF (5.5 MB)

More Alcatel Manuals


View all Alcatel manuals →