The Signet 463X Chlorine Analyzer System is a compact, integrated chlorine panel system designed to measure free chlorine and chlorine dioxide in drinking water and clean freshwater treatment applications. This manual covers the 4630 Free Chlorine Analyzer System and the 4632 Chlorine Dioxide Analyzer System.
This manual provides instructions for installing, calibrating, and maintaining the Signet 463X chlorine analyzer system, designed for measuring free chlorine and chlorine dioxide in potable and clean water applications. It includes detailed sections on sensor preparation, flow element assembly, transmitter operation, and troubleshooting. The system features an integrated flow element, NEMA 4X enclosure, and automatic pH/temperature compensation.
Technical specifications cover sensor compatibility, flow range (30.24–45.36 LPH), electrical requirements (100–240 VAC), and environmental operating conditions (0–45°C). Safety precautions and warranty information are also included. This manual is written in Spanish.
| Sensor Compatibility | 3-2630-1, 3-2630-2, 3-2630-3 for free chlorine; 3-2632-1 for chlorine dioxide; 3-2724-00 for pH |
|---|---|
| Flow Range | 30.24 to 45.36 LPH (8 to 12 gph: US) |
| Input Voltage | 100 to 240 VAC nominal (85 to 264 VAC), 50 to 60 Hz (47 to 63 Hz) |
| Operating Temperature | 0 to 45 °C (32 to 113 °F) |
| Enclosure Rating | NEMA 4X (with output wire glands sealed) |
Compatible sensors include 3-2630-1, 3-2630-2, 3-2630-3 for free chlorine; 3-2632-1 for chlorine dioxide; and 3-2724-00 for pH.
The flow range is 30.24 to 45.36 LPH (8 to 12 gph: US).
The system operates between 0 to 45 °C (32 to 113 °F).
The enclosure is rated NEMA 4X and is moisture-resistant with sealed output wire glands.