Signet 2610 Process Optical Dissolved Oxygen Sensor Operator's Manual

This document is the user's manual for the Signet 2610 Process Optical Dissolved Oxygen Sensor. The sensor features the latest optical technology for DO measurement, and can deliver accurate dissolved oxygen (DO) data across a wide measuring range while reducing maintenance costs.


Signet 2610 Process Optical Dissolved Oxygen Sensor Operator's Manual - cover page
Brand
Signet
Category
Other Electronics
Document type
User Manual
Language
English
Pages
8
File format
PDF
File size
605 KB
Published
20 October, 2023
Updated
20 October, 2023
MD5 checksum
75AED717093676F6DEB8DA32EA7A914A

About this manual

The Signet 2610 Process Optical Dissolved Oxygen Sensor is a rugged, reliable device designed to deliver accurate dissolved oxygen data across a wide measuring range. It features optical technology and comes with three standard outputs: Digital (S 3L), 4 to 20 mA, and ModBus RS485. The sensor body includes a removable nose cone, a 10-meter cable, an optical DO sensing cap, and a titanium temperature sensor.

What's inside

  • Warranty Information
  • Safety Information
  • Chemical Compatibility
  • Care and Maintenance
  • Specifications
  • Dimensions
  • Unpacking the Sensor
  • Calibration
  • Wiring
  • Sensor Installation

Specifications

Sensor typeLuminescent dissolved oxygen sensor
Communications optionsDigital (S 3L), Modbus (RS485), 4 to 20 mA
Cable length32.8 ft/10 meters
Internal mounting thread1¼ NPT
Salinity range0 to 42 PSU
pH range2 to 10 pH
Barometric range507 to 1115 mbar
Maximum pressure300 psi
Range0 to 20 mg/L concentration, 0 to 200% saturation
Accuracy (DO)±0.1 mg/L, 0 to 8 mg/L, ±0.2 mg/L, 8 to 20 mg/L
Response timeCap T90: 30 sec, Cap T95: 37 sec @ 25 °C
Resolution0.01 mg/L
Operating temperature0 °C to 50 °C (32 °F to 122 °F)
IP ratingIP-67 with cap off, IP-68 with cap installed
Warranty Sensor3 years from date of manufacture

Frequently asked questions

Is calibration required for the Signet 2610 sensor?

No, calibration is not required as the unit measures within 2% of reading for the life of the sensor cap.

How often should the sensor cap be replaced?

The sensor cap has a 2-year life after the instrument takes its first reading.

What are the wetted materials of the sensor?

The wetted materials are ABS, Titanium, and FKM.

Can the S3L and Modbus connectors be used simultaneously?

Yes, the S3L and Modbus connectors can be used simultaneously with the 4 to 20 mA output.

Download PDF (605 KB)

More Signet Manuals