Signet 4-3000-5 Remote AgRotor Flow Sensor Assembly User Manual

The manual introduces the Signet 4-3000-5 AgRotor Flow Sensor Assembly, which can be used to measure water flow in irrigation piping systems, has the characteristics of reliability and accuracy, and does not require recalibration after routine maintenance. It can be mounted at any angle for versatile installation. It has an innovative rotor design, one moving part, snap-on rotor replacement (no special tools required), and reminds the user to pay attention to the risk of electrostatic discharge and electric shock.


Signet 4-3000-5 Remote AgRotor Flow Sensor Assembly User Manual - cover page
Brand
Signet
Document type
User Manual
Language
English
Pages
4
File format
PDF
File size
1.1 MB
Published
20 October, 2023
Updated
20 October, 2023
MD5 checksum
CCC261A1834C01F09812AF2E38A66981

About this manual

This manual provides instructions for installing and maintaining the Signet 4-3000-5 Remote AgRotor Flow Sensor Assembly in irrigation piping systems. It includes safety warnings, mounting procedures, wiring details, and rotor replacement steps. The sensor uses an electromagnetic signal from the rotor tip to measure flow without recalibration after maintenance. It can be mounted at any angle and features a snap-on rotor design for easy replacement without special tools.

The manual also covers specifications such as pipe size compatibility, accuracy, cable length, and electrical requirements. Maintenance guidelines recommend checking for wear at least once per season and include steps for depressurizing the system before servicing. Replacement parts and accessories are listed with part numbers for easy ordering.

What's inside

  • Safety Information
  • Installation
  • Wiring
  • Maintenance and Repair
  • Rotor Replacement
  • Specifications
  • Ordering Information

Specifications

Typical Pipe Size and Flow Rates6 in. pipe … 88 to 1,761 gpm; 8 in. pipe … 157 to 3,132 gpm; 10 in. pipe … 245 to 4,893 gpm; 12 in. pipe … 352 to 7,046 gpm
Repeatability±0.5% of maximum range
Accuracy±2% of reading (under reference conditions)
Cable Length25 ft supplied, can be increased to 100 ft by customer
Cable Type2-conductor shielded twisted pair, 14-22 AWG
Rotor Diameter5 in.
Sensor Body MaterialPC/PBT Alloy
Rotor Assembly MaterialPC/PBT Alloy and Acetal
Sensor Flange MaterialA-36 Cold Roll Steel with Galvanized Coating
Sensor Gasket MaterialNeoprene Rubber
Rotor Bearing and Shaft MaterialCeramic
Rotor Face Seal Material316 SS
Max. Pressure/Temperature Ratings150 psi max @ 68°F; 32 psi max @ 140°F
Electrical Supply Voltage3.5 to 27 VDC
Supply Current<20 μA with sensor powered with 3.5-5 VDC; <3mA with sensor powered with 5-27 VDC
Signal OutputOpen collector transistor, sink, max. pullup voltage 27 VDC (transistor normally off)
Max. Open-Collector Current Out20 mA
Output Pulse Width3 msec. nominal
Standards and ApprovalsCE, ISO 9001, ISO 14001, OHSAS 18001

Frequently asked questions

How do I install the sensor?

Install the sensor by drilling a 2.5-inch hole in the pipe, placing the flanged fitting half over the hole, and tightening bolts to no more than 65 foot-pounds.

How do I replace the rotor?

Insert a screwdriver or coin under the rotor retaining clip to release the rotor assembly, then gently separate the sensor halves. Reinstall in reverse order, ensuring the retainer clip clicks into position.

What is the accuracy of the sensor?

The accuracy is ±2% of reading under reference conditions.

What is the maximum pressure rating?

The maximum pressure rating is 150 psi at 68°F, or 32 psi at 140°F.

Download PDF (1.1 MB)