Signet 2760 DryLoc pH/ORP Preamplifier User Manual

The Signet 2760 DryLoc® pH/ORP Preamplifiers with DryLoc® connectors are part of Signet's growing line of sensors for industrial fluid measurement and control. There are two styles of preamplifiers, and each is available with NPT or ISO threads at the cable outlet. They all connect to four different families of pH and ORP electrodes to supply an amplified mV (analog) signal to Signet instruments.


Signet 2760 DryLoc pH/ORP Preamplifier User Manual - cover page
Brand
Signet
Category
Industrial Robot
Document type
User Manual
Language
English
Pages
8
File format
PDF
File size
1.9 MB
Published
20 October, 2023
Updated
20 October, 2023
MD5 checksum
1304C45639B24DE0E42A10D8E822C405

About this manual

This manual provides installation instructions for Signet 2760 DryLoc pH/ORP Preamplifiers, including in-line, wet-tap, and submersible setups. It details wiring to Signet instruments and compatible electrodes. Safety precautions and environmental specs are included.

Technical specifications cover electrical parameters, enclosure ratings, materials, and operating/storage temperatures. The document also includes diagrams for assembly and notes on chemical compatibility and warranty conditions.

What's inside

  • Installation
  • Wiring to Signet Instruments
  • Wiring to Other Equipment
  • Specifications
  • Chemical Compatibility
  • Safety Information
  • Warranty Information
  • Ordering Information

Specifications

Response Time (pH)< 6 sec. for 95% of change
Temperature, τ20 sec.
Maximum Offset± 7 mV
Input Impedance, Z> 1011 ℧
Supply Voltage6 to 12 VDC
Current Consumption< 1 mA
Enclosure Rating (In-line)NEMA 6P/IP68
Cable Length4.6 m (15 ft)
Maximum Cable Length122 m (400 ft)
Shipping Weight0.64 kg (1.41 lb)
Operating Temperature0 °C to 85 °C (32 °F to 185 °F)
Storage Temperature–20 °C to 85 °C (–4 °F to 185 °F)
Relative Humidity0 to 95%, non-condensing (without electrode connected)

Frequently asked questions

What types of electrodes are compatible with the 2760 DryLoc Preamplifier?

Compatible electrodes include 2724-2726, 2756-WTP, 2757-WTP, 2764-2767, and 2774-2777 series.

Which Signet instruments can the 2760 Preamplifier connect to?

It can connect to Signet 5700 and 8750 series instruments.

What is the maximum cable length for the 2760 Preamplifier?

The maximum cable length is 122 m (400 ft).

What safety precautions should be taken during installation?

Depressurize and vent the system, wear safety goggles or faceshield, and do not exceed temperature/pressure limits.

Download PDF (1.9 MB)