Druck DPI 705/DPI 705 IS Series Digital Pressure Indicator User Manual

Instruction manual is a book which is used to guide users to correctly use the product. This manual is suitable for Druck DPI 705/DPI 705 IS series digital pressure gauge.


Druck DPI 705/DPI 705 IS Series Digital Pressure Indicator User Manual - cover page
Brand
Druck
Category
pressure transmitters
Document type
User Manual
Language
English
Pages
42
File format
PDF
File size
831 KB
Published
17 October, 2023
Updated
17 October, 2023
MD5 checksum
F1A778FCA3ABF1BE79678612425AA3C4

About this manual

The document serves as a comprehensive user guide for the Druck DPI 705 and DPI 705 IS Series Digital Pressure Indicators. It covers general operation features such as leak testing, zeroing, alarm settings, and display symbols. The manual also details specific requirements for using Intrinsically Safe (IS) versions in hazardous areas, including battery specifications and installation warnings.

What's inside

  • Specification
  • Safety
  • Operation
  • Intrinsically Safe Pressure Indicators
  • CSA Certification Markings
  • Maintenance
  • Cleaning

Specifications

Accuracy (Combined non-linearity, hysteresis and repeatability)±0.1% full-scale (FS)
Operating range-10°C to 50°C
Storage range-20°C to 60°C
Maximum safe working pressure2 x full-scale
Pressure connector6 mm o/d and 4 mm i/d hose, G1/8 or 1/8"NPT female thread
Electrical Power Supply (non-certified units)3 x 1.5 V alkaline size AA

Frequently asked questions

What is the procedure for zeroing the instrument?

Open all pressure ports to atmospheric pressure and press the TARE keys together.

What batteries are required for Intrinsically Safe units?

Use only 3 x LR6 (AA) Duracell PROCELL, PLUS, ENERGIZER ULTIMATE or GP SUPERALKALINE.