MycoGuardTM Mycoplasma Bioluminescent Detection Kit is an efficient and selective biochemical detection kit that can quickly and easily detect the presence of mycoplasma in cell culture processes.
The MycoGuardTM Mycoplasma Bioluminescent Detection Kit is designed for rapid and sensitive detection of mycoplasma contamination in cell cultures. It operates by measuring the activity of specific enzymes released from lysed mycoplasma, which catalyze a reaction converting ADP to ATP. The resulting increase in ATP concentration generates a measurable bioluminescent signal. This manual is written in Chinese.
The manual recommends centrifuging the cell culture at 1500rpm (200g) for 5 minutes to remove residual cells before detection.
After adding 50µL of Myco-Sub, the mixture must be mixed and incubated for 10 minutes before reading value B.
A ratio between 0.9 and 1.2 is considered suspicious and suggests the test should be repeated after 24 hours.