Hamamatsu L2D2 Deuterium Lamps Selection Guide

The document describes the best electrode technology supported by the best light source, and provides detailed information on the features and performance of L2D2 lamps.


Hamamatsu L2D2 Deuterium Lamps Selection Guide - cover page
Brand
Hamamatsu
Category
lighting
Document type
Product Leaflet
Language
English
Pages
7
File format
PDF
File size
1.1 MB
Published
10 August, 2012
Updated
29 September, 2023
Model
L2D2
MD5 checksum
789492C6E7417926EA90472F4D990C1F

About this manual

The document presents the Hamamatsu L2D2 deuterium lamp line, including the L2-2000 and L2-4000 series. It states a 4000-hour guaranteed life for L2-4000 lamps, 1.3 times higher light output for L2-2000 lamps, and output stability of 0.05 %p-p fluctuation and ±0.3 %/h drift.

It also lists applications such as HPLC, atomic absorption spectrophotometers, thin layer chromatography, UV-VIS spectrophotometers, CE, SOx/NOx analyzers and film thickness measurement. Specification tables, dimensional outlines, connection details and see-through type information are included.

What's inside

  • Applications
  • Long Life
  • High Light Output
  • High Stability
  • Excellent Temperature Characteristics
  • Small Intensity Variations
  • See-through Type
  • Selection Guide
  • Specifications for L2D2 Lamps
  • Dimensional Outlines

Specifications

Guaranteed life (L2-4000 series)4000 hours
Light output (L2-2000 series)1.3 times higher than conventional lamps
Light output fluctuation0.05 %p-p
Drift±0.3 %/h
Recommended operating voltage3.5 V ± 0.5 V
Recommended discharge starting voltage500 V to 600 V; maximum 650 V
Aperture diameter0.5 or 1.0 mm

Frequently asked questions

What is the guaranteed operating life of the L2-4000 series lamps?

The L2-4000 series lamps have an operating life of 4000 hours, stated as four times longer than conventional lamps.

How much higher is the light output of the L2-2000 series?

The L2-2000 series produces 1.3 times higher light output than conventional lamps.

What starting voltage is recommended for reliable lamp lighting?

An application of 500 V to 600 V is recommended; the maximum rated voltage is 650 V.

Are L2D2 lamps always direct replacements for conventional lamps?

No. The document says L2D2 lamps do not always have a direct replacement from a dimensional outline point of view.