Hanna Instruments HI96752 Calcium and Magnesium Meter Instruction Manual

The HI 96752 Calcium and Magnesium High Range ISM is a portable ion-selective conductivity meter for measuring calcium and magnesium in water. The instrument employs advanced microprocessor technology and a full automatic ion-selective electrode to provide fast, accurate, and reliable measurement results.


Hanna Instruments HI96752 Calcium and Magnesium Meter Instruction Manual - cover page
Brand
Hanna Instruments
Category
Other Electronics
Document type
User Manual
Language
English
Pages
2
File format
PDF
File size
695 KB
Published
10 July, 2012
Updated
30 September, 2023
MD5 checksum
5DFB2A7DA312E55AB3DDCAD68A4B20B4

About this manual

The Hanna Instruments HI96752 is a high-range ISM (Indicating Spectrophotometric) instrument designed to measure Calcium and Magnesium concentrations. It utilizes the Orolate method for Calcium and the Calmogite method for Magnesium, featuring an LED light source and a silicon photocell detector with a narrow band interference filter at 466 nm.

What's inside

  • Preliminary examination
  • Technical specifications
  • Functional description
  • Errors and warnings
  • Measurement procedure
  • Interferences
  • Validation procedures
  • Calibration procedures

Specifications

Range Magnesium0 to 150 mg/l
Resolution1 mg/l
Accuracy Calcium+/- 10% of reading at 25°C
Accuracy Magnesium+/- 3 mg/l +/- 3% of reading at 25°C
Typical EMC Dev.1 mg/l
Light SourceLED
DetectorSilicon Photocell with narrow band interference filter @ 466 nm
Environment Temperature0 to 50°C (82 to 122°F)
Battery Type1 x 9 volt

Frequently asked questions

What methods are used for Calcium and Magnesium measurement?

The instrument uses the Orolate method adaptation for Calcium and the Calmogite method adaptation for Magnesium.

What are the accuracy specifications at 25°C?

For Calcium, accuracy is +/- 10% of reading; for Magnesium, it is +/- 3 mg/l or +/- 3% of reading.

What happens if the battery is low?

A 'Battery low' warning appears, indicating the battery must be replaced soon. If dead, normal operation is interrupted.

Can I use standard solutions other than Hanna CAL CHECK for calibration?

No, you should not validate or calibrate with other standards as it will result in erroneous readings.

Download PDF (695 KB)

More Hanna Instruments Manuals


View all Hanna Instruments manuals →