Hanna Instruments HI96771 Free Chlorine & Chlorine Ultra High Range ISM Instruction Manual

This document introduces the use of HI 95771A-0 / HI 95771B-0 Chlorine UHR (ultra high sensitivity)water quality test meter, including instrument power on, sample testing, measurement results and other steps.


Hanna Instruments HI96771 Free Chlorine & Chlorine Ultra High Range ISM Instruction Manual - cover page
Brand
Hanna Instruments
Category
Other Electronics
Document type
User Manual
Language
English
Pages
2
File format
PDF
File size
675 KB
Published
10 July, 2012
Updated
30 September, 2023
MD5 checksum
BC2B2D5A1F8A084DFE24EF0111D813A3

About this manual

This document is the instruction manual for the Hanna Instruments HI96771 Free Chlorine & Chlorine Ultra High Range ISM (In Situ Meter). It provides comprehensive guidance on the preliminary examination of the instrument, including checking for damage and verifying the included accessories such as sample cuvettes, caps, and batteries. The manual details the technical specifications, including measurement ranges, accuracy, resolution, and environmental requirements for both Chlorine Low Range (LR) and Chlorine Ultra High Range (UHR) modes.

What's inside

  • Preliminary examination
  • Technical specifications
  • Functional description
  • Errors and warnings
  • Measurement procedure
  • Interferences
  • Validation and Calibration procedures
  • GLP mode

Specifications

Range (Chlorine LR)0.00 to 5.00 mg/L
Range (Chlorine UHR)0 to 200 mg/l
Accuracy (Chlorine LR)0.09 mg/L +3% of reading
Accuracy (Chlorine UHR)3 mg/L +3% of reading
Resolution (Chlorine LR)0.01 mg/L (0.00-3.50 mg/l)
Resolution (Chlorine UHR)1 mg/L
Light SourceTungsten lamp
Light DetectorSilicon Photocell
Battery Typex 9 volt
Dimensions192 x 104 x 69 mm

Frequently asked questions

What is the accuracy for the Chlorine Low Range?

The accuracy is 0.09 mg/L +3% of reading at 25°C.

What is the accuracy for the Chlorine Ultra High Range?

The accuracy is 3 mg/L +3% of reading at 25°C.

What happens if the sample concentration is beyond the programmed range?

An Over Range condition occurs, indicated by a flashing value of the maximum concentration.

What should be done if the standard reading is less than expected during calibration?

The error message 'Standard Low' will appear on the display.

Download PDF (675 KB)

More Hanna Instruments Manuals


View all Hanna Instruments manuals →