instruction manual
The HI 96701C is an auto-diagnostic portable microprocessor meter designed to measure free chlorine content in water samples. It features an advanced optical system with a tungsten lamp and narrow band interference filter, ensuring accurate and repeatable readings. The device includes a CAL CHECK validation function to verify performance using NIST traceable standards.
| Range | 0.00 to 5.00 mg/L |
|---|---|
| Resolution | 0.01 mg/L from 0.00 to 3.50 mg/L; 0.10 mg/L above 3.50 mg/L |
| Accuracy | +/-0.03 mg/L +3% of reading @ 25°C |
| Light Source | Tungsten lamp |
| Light Detector | Silicon Photocell with narrow band interference filter @ 525nm |
| Environment | 0 to 50°C (32 to 122°F); max 95% RH non-condensing |
| Battery Type | 1x9 volt |
| Dimensions | 192 x 104 x 69 mm (7.6 x 4.1 x 2.7") |
| Weight | 360 g (12.7 oz) |
The meter measures free chlorine content in the 0.00 to 5.00 mg/L (ppm) range.
Use the CAL CHECK function with exclusive HANNA ready-made, NIST traceable standards to verify performance.
The meter has an auto-shut off feature that turns off the instrument after 10 minutes of non-use in measurement mode or after 1 hour in calibration mode.
The required reagent is HI 93701-0 DPD Powder Reagent.