Hanna Instruments HI 96713C Phosphate Low Range ISM Instruction Manual

The HI 96713C Phosphate Low Range ISM is an electrochemical analyzer manufactured by Hanna Instruments. It is mainly used to measure the phosphate content.


Hanna Instruments HI 96713C Phosphate Low Range ISM Instruction Manual - cover page
Brand
Hanna Instruments
Category
Analyzers
Document type
User Manual
Language
English
Pages
24
File format
PDF
File size
1.2 MB
Published
10 July, 2012
Updated
30 September, 2023
MD5 checksum
4367AD0F6736DD8F42DDC2B1B543C167

About this manual

The HI 96713C is an auto-diagnostic portable microprocessor meter designed to measure phosphate content in water samples within the range of 0.00 to 2.50 mg/L (ppm). It utilizes an advanced optical system based on a special Tungsten Lamp and a narrow band interference filter to ensure accurate and repeatable readings. The instrument features a powerful CAL CHECK validation function and is splash waterproof for field applications.

What's inside

  • Preliminary Examination
  • General Description
  • Specifications
  • Precision and Accuracy
  • Principle of Operation
  • Functional Description
  • Errors and Warnings
  • Startup
  • Measurement Procedure
  • Calibration Procedure

Specifications

Range0.00 to 2.50 mg/L
Resolution0.01 mg/L
Accuracy±0.04 mg/L ± 4% of reading @ 25°C
Light SourceTungsten lamp
Light DetectorSilicon Photocell with narrow band interference filter @ 610 nm
MethodAdaptation of the ascorbic acid method
Environment0 to 50°C (32 to 122°F); max 95% RH non-condensing
Battery1 x 9 volt
Dimensions192 x 104 x 69 mm (7.6 x 4.1 x 2.7")
Weight360 g (12.7 oz.)

Frequently asked questions

What is the measurement range of the HI 96713C?

The meter measures phosphate content in the range of 0.00 to 2.50 mg/L (ppm).

What method is used for the measurement?

The method is an adaptation of the ascorbic acid method.

How long does the auto-shut off feature last?

The instrument turns off after 10 minutes of non-use in measurement mode or after 1 hour in calibration mode.

What does the CAL CHECK function do?

It allows the user to validate the good performance of the instrument at any time using NIST traceable standards.

Download PDF (1.2 MB)

More Hanna Instruments Manuals


View all Hanna Instruments manuals →