Hanna HI 96711C Free and Total Chlorine ISM Instruction Manual

This manual will provide you with the necessary information for the correct use of the instrument. Please read it carefully before using the meter. If you need additional technical information, do not hesitate to e-mail us at [email protected].


Hanna HI 96711C Free and Total Chlorine ISM Instruction Manual - cover page
Brand
Hanna Instruments
Category
Other Electronics
Document type
User Manual
Language
English
Pages
15
File format
PDF
File size
1.3 MB
Published
10 July, 2012
Updated
30 September, 2023
MD5 checksum
45ADAED74F5D36937C377C5FDE97FF6F

About this manual

The Hanna HI 96711C is an auto-diagnostic portable microprocessor meter designed to measure free and total chlorine content in water samples. It utilizes an advanced optical system based on a special tungsten lamp and a narrow band interference filter to ensure accurate and repeatable readings. The instrument is splash waterproof and features a positive-locking system for the cuvette to maintain consistent measurement conditions.

What's inside

  • General Description
  • Specifications and Accuracy
  • Errors and Warnings
  • General Tips for Accurate Measurement
  • Startup
  • Range Selection
  • Measurement Procedure
  • Validation Procedure
  • Calibration Procedure
  • Battery Management

Specifications

Range0.00 to 5.00 mg/L
Resolution0.01 mg/L from 0.00 to 3.50 mg/L; 0.10 mg/L above 3.50 mg/L
Accuracy±0.03 mg/L ±3% of reading @ 25°C
Light SourceTungsten lamp
Light DetectorSilicon Photocell with narrow band interference filter @ 525nm
Environment0 to 50°C (32 to 122°F); max 95% RH non-condensing
Battery1 x 9 volt
Dimensions192 x 104 x 69 mm (7.6 x 4.1 x 2.7")
Weight360 g (12.7 oz.)

Frequently asked questions

What is the measurement range for free and total chlorine?

The meter measures free and total chlorine in the 0.00 to 5.00 mg/L (ppm) range.

What happens if the battery is low?

The instrument displays a 'Battery low' message, indicating the battery must be replaced soon.

How do I enter calibration mode?

Press and hold the CAL CHECK key for three seconds until the 'CAL' tag appears.

What method does the instrument adapt?

The method is an adaptation of the USEPA Method 330.5 for wastewater and Standard Method 4500-Cl G for drinking water.