Calibrating the Response of the Varian ELSD Application Note

Varian ELSD Advantage Statement: The high sensitivity of the Varian Evaporative Light Scattering Detector is evident from the large response it delivers, even with low column loading. This note describes the calibration of the Varian ELSD response with a GPC application.


Calibrating the Response of the Varian ELSD Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
1
File format
PDF
File size
118 KB
Published
05 July, 2012
Updated
30 September, 2023
MD5 checksum
4881877C3F79D99D89F34AC8AFE8E164

About this manual

This note details the calibration of the Varian Evaporative Light Scattering Detector (ELSD) response within a Gel Permeation Chromatography (GPC) application. The procedure involves determining the detector's response for specific calibration solutions to ensure accurate measurements.

The test utilized a polystyrene narrow standard (Mp 30,300) eluting through a PLgel MIXED-D column with a fixed injection volume of 100 µL. A series of calibration solutions ranging from 1 to 40 ng of column loading were prepared by diluting a stock solution. The results demonstrated that the ELSD provides a linear response to column loading, even at low concentrations.

What's inside

  • Calibration procedure
  • GPC application
  • Polystyrene standard
  • Column loading
  • Detector response
  • Linear fit analysis