Agilent PLgel GPC/SEC Technical Overview: Effect of Flow Rate on Oligomer Separation

This white paper introduces the oligomer separation effect of Agilent PLgel 5 µm columns at different flow rates


Agilent PLgel GPC/SEC Technical Overview: Effect of Flow Rate on Oligomer Separation - cover page
Brand
Agilent
Category
Other
Document type
Product Leaflet
Language
English
Pages
2
File format
PDF
File size
46 KB
Published
05 July, 2012
Updated
30 September, 2023
MD5 checksum
D427EB009FC024F79B692CA75E172C62

About this manual

This document provides a technical overview of the effect of flow rate on oligomer separation in gel permeation chromatography (GPC). It illustrates that resolution improves as the flow rate is reduced, although this results in increased run time. The demonstration utilizes an Agilent PLgel 5 µm column with three different flow rates.

What's inside

  • Introduction
  • Effect of Flow Rate
  • Separation of Polystyrene Standards
  • Conditions
  • Calibrants
  • Columns
  • Eluent
  • Detector
  • System
  • Agilent Literature Library

Specifications

Column TypeAgilent PLgel 5 µm
Column Part NumberPL1110-6520
Column Dimensions300 x 7.5 mm
EluentTHF
SystemAgilent PL-GPC 50
CalibrantAgilent Polystyrene Standard 480

Frequently asked questions

How does flow rate affect oligomer separation resolution?

The resolution of oligomers improves as the flow rate is reduced.

What is the trade-off of reducing flow rate?

The run time is obviously increased as a result of reducing the flow rate.