Agilent PLgel 100A Columns Technical Overview

This note illustrates the effect of particle size and injection volume on the efficiency of Agilent PLgel 100Å, 300 × 7.5 mm columns.


Agilent PLgel 100A Columns Technical Overview - cover page
Brand
Agilent
Document type
Product Leaflet
Language
English
Pages
2
File format
PDF
File size
64 KB
Published
05 July, 2012
Updated
01 October, 2023
MD5 checksum
AE99B374CAB64F1A5D343CD6366244AC

About this manual

The text discusses two factors affecting column efficiency and resolution: the particle size of packing materials and system dead volume. Smaller particles lead to higher efficiencies, but minimizing extra-column dispersion is necessary for optimal performance with small particle sizes.

It illustrates these concepts using Agilent PLgel 3 pm, 5 pm, and 10 pm 100A columns tested with polystyrene 580 samples. The document demonstrates that decreasing injection volume from 100 pL to 20 pL significantly improves resolution.

What's inside

  • Particle Size
  • Injection Volume
  • Column Efficiency
  • Resolution
  • Dead Volume

Frequently asked questions

What are the dimensions of the Agilent PLgel 100A columns discussed in this overview?

The document specifies that the Agilent PLgel 100A columns have dimensions of 300 x 7.5 mm.

What flow rate was used for the chromatograms shown in Figure 2?

The chromatograms were obtained using a flow rate of 1.0 mL/min.

What is the effective molecular weight range for the Agilent PLgel 3 um 100A column?

The effective MW range for this column extends up to 4,000.