Agilent 710 Series ICP-OES spectrometers are manufactured according to a quality management system certified to ISO 9001. The 710 and 715 ICP-OES systems are compact, bench-mounted simultaneous Inductively Coupled Plasma Optical Emission Spectrometers (ICP-OES) with full PC control and compatible accessories. They feature an innovative megapixel CCD detector designed specifically for ICP-OES and provide complete wavelength coverage from 177–785 nm. Agilent ICP-OES systems are available in either axially (710) or radially (715) viewed configurations and include a free-running, air-cooled 40 MHz RF generator, purged echelle polychromator, and full PC control of plasma viewing position and plasma gas flows. The fully web-integrated ICP Expert II software uses Agilent’s worksheet concept for ease of use, rapid operator training, and commonality with Agilent Atomic Absorption products.
The Agilent 710 and 715 Series ICP-OES systems are compact, bench-mounted simultaneous Inductively Coupled Plasma Optical Emission Spectrometers designed for accurate, robust, and reliable analysis. They feature an innovative megapixel CCD detector and full PC control via ICP Expert II software, offering complete wavelength coverage from 177-785 nm. The systems are available in axially viewed (710) or radially viewed (715) configurations to suit specific application needs, such as high sensitivity analyses or difficult matrices like oils and organic solvents.
| Wavelength coverage | 177-785 nm |
|---|---|
| Detector type | Megapixel CCD |
| RF generator frequency | 40 MHz |
| Power output range | 700-1500 W |
| Warm-up time | <10 mins |
| Stray light limit | <2.0 mg/L |
| Signal stability | <1% RSD over 8 hours |
| Resolution (FWHM) for As | <9 pm |
The 710 is an axially viewed system ideal for high sensitivity, while the 715 is a radially viewed system ideal for difficult applications like oils and organic solvents.
The warm-up time from standby mode is less than 10 minutes from plasma ignition.
The system is typically stable to less than 1% RSD over 8 hours without internal standardization or drift correction.