Varian 310-MS LC/MS Product Specifications

Varian 310-MS LC/MS is a high performance mass spectrometer, suitable for applications requiring quantitation in heavy matrices, such as pharmaceuticals, toxicology, and environmental.


Varian 310-MS LC/MS Product Specifications - cover page
Brand
Agilent
Category
Other
Document type
Product Leaflet
Language
English
Pages
2
File format
PDF
File size
117 KB
Published
04 July, 2012
Updated
30 September, 2023
Model
310-MS
MD5 checksum
B83B02CF3D4A9ED90D2FB00C90ED7BA9

About this manual

The Varian 310-MS LC/MS is a system designed for pharmaceutical, toxicological, and environmental applications requiring quantitation in heavy matrices. It features a flexible and affordable design with a unique MS analyzer that delivers superior signal-to-noise performance.

The document outlines technical specifications for the mass analyzer, including a mass range of 10 to 2000 Da and scan rates up to 6000 Da/sec. It details the Atmosphere Pressure Ionization (API) and Electrospray Interface (ESI) capabilities, along with a list of compatible chromatography options such as AutoSamplers and Solvent Delivery Modules.

What's inside

  • Specifications
  • Atmosphere Pressure Ionization Interface
  • Electrospray Interface
  • Mass Filters
  • Collision Cell
  • Vacuum System
  • Utilities and Environment
  • User Upgradeable Options
  • Chromatography Options
  • Performance Specifications

Specifications

Analyzer mass range10 to 2000 Da
Scan rates500 to 6000 Da/sec
Resolutionunit mass over the entire range
Mass axis stability+0.1 Da over 24 hours
Turbomolecular pumpdual stage, 310/400 L/sec, air cooled
Power requirements for the MS100-240 Vac, 50/60 Hz +3 Hz, 1.0 kW

Frequently asked questions

What applications is the Varian 310-MS LC/MS suitable for?

It is suitable for pharmaceutical, toxicological, and environmental applications requiring quantitation in a heavy matrix.

What is the mass range of the analyzer?

The analyzer mass range is 10 to 2000 Da.

Can the system switch between ESI and APCI?

Yes, the APCI option uses the same hinge/clip mounting system as the ESI chamber and can be changed in a few minutes.