Agilent 720/730 Series ICP-OES Specifications Manual

Agilent 720/730 Series ICP-OES spectrometer is a compact bench-mounted simultaneous Inductively Coupled Plasma Optical Emission Spectrometers (ICP-OES) manufactured by Agilent, featuring a custom-designed CCD detector with unique image mapping (I-MAP) technology, providing complete wavelength coverage from 167-785 nm.


Agilent 720/730 Series ICP-OES Specifications Manual - cover page
Brand
Agilent
Document type
Product Leaflet
Language
English
Pages
8
File format
PDF
File size
105 KB
Published
04 July, 2012
Updated
30 September, 2023
MD5 checksum
C3C15E0A7A5E0A9260867146C61E8785

About this manual

The 720/730 Series are compact, bench‑mounted ICP‑OES spectrometers with full PC control, offering radial (725/735) or axial (720/730) viewing configurations. They feature a custom‑designed CCD detector with I‑MAP technology, a 40 MHz RF generator, and software‑controlled gas flows.

The manual lists analytical performance (warm‑up time, stray‑light, signal stability), resolution, detection limits for numerous elements, and a range of accessories such as autosamplers, diluters, and vapor generators. Installation requirements and support policies are also included.

What's inside

  • Introduction
  • Design overview
  • Sample introduction
  • Instrument hardware
  • Instrument software
  • Analytical performance
  • Accessories
  • Installation requirements
  • Customer support policies

Specifications

Wavelength coverage167–785 nm
RF generator frequency40 MHz
RF power output range700–1500 W
Power output stability<0.1 %
Signal stability<1 % RSD over 8 h
Plasma gas flow range0–22.5 L/min
Auxiliary gas flow range0–2.25 L/min
CCD detector cooling temperature-35 °C

Frequently asked questions

What wavelength range does the CCD detector cover?

The detector provides full wavelength coverage from 167 nm to 785 nm.

What is the RF generator frequency and power range?

The RF generator operates at 40 MHz with power output adjustable between 700 W and 1500 W.

What is the detection limit for silver (Ag) in axial mode?

The 3‑sigma detection limit for Ag at 328.068 nm in axial mode is 0.3 µg/L.