The 709-DS Dissolution Apparatus uses the same platform as the 708-DS, ensuring productivity, reproducibility and ease of use combined with the convenience of a bath-free design. The 709-DS reduces media heating time with Direct Vessel Heating (DVH) capabilities while insulating the vessel from environmental factors and eliminating water bath cleaning. The exclusive design of the DVH ensures that the temperature inside the vessel is constant, eliminating temperature drift. The 709-DS uses an advanced electronic control system to precisely control the temperature and speed of the solution, ensuring the accuracy of the test results. The 709-DS is equipped with an intuitive touch screen for easy operation and setup.


Agilent 709-DS Dissolution Apparatus Product Information - cover page
Brand
Agilent
Document type
Product Leaflet
Language
English
Pages
1
File format
PDF
File size
329 KB
Published
04 July, 2012
Updated
30 September, 2023
Model
709-DS
MD5 checksum
C13C267ACCDFAFEB47427939ACC77AA6

About this manual

The Agilent 709-DS Dissolution Apparatus is designed to ensure productivity, reproducibility, and ease of use. It utilizes a unique bath-free design with Direct Vessel Heating (DVH) technology to reduce media heating time while insulating the vessel from environmental factors.

Key features include an exclusive DVH TruAlign vessel incorporating heat-sensing technology in the vessel wall and a clear vessel coating for visibility and durability. The apparatus is compatible with optional features such as the Dosage Delivery Module (DDM), Autosampling, handheld temperature probes, and printers.

What's inside

  • Specifications
  • Direct Vessel Heating
  • DVH TruAlign Vessel
  • Operating Requirements
  • Optional Features

Specifications

Evaporative lossLess than 1%
Ambient temperature range16 to 36 °C
Humidity limitNot more than 80% RH
Voltage operating range90-250 V, 50-60 Hz
Temperature RangeAmbient +5 to 55°C
Spindle Speed Range10-250 RPM
Speed Accuracy+1% over 25 RPM
Weight54.4 kg (120 Ib)

Frequently asked questions

What is the main advantage of the 709-DS design?

It features a bath-free design that reduces media heating time and eliminates water bath cleaning.

What optional features are available for the apparatus?

Available options include the Dosage Delivery Module, Autosampling, handheld temperature probe, and printer.