Agilent 240 MS Ion Trap Mass Spectrometer Data Sheet

Agilent 240 Ion Trap Mass Spectrometer


Agilent 240 MS Ion Trap Mass Spectrometer Data Sheet - cover page
Brand
Agilent
Document type
Datasheet
Language
English
Pages
4
File format
PDF
File size
55 KB
Published
04 July, 2012
Updated
30 September, 2023
Model
240-MS
MD5 checksum
95DDC5744F7B888E133258BF4F8F5960

About this manual

The Agilent 240 MS is an ion trap mass spectrometer designed to deliver exceptional performance when combined with the Agilent 7890A GC. It supports both internal and external ionization modes, offering MS/MS capabilities and positive or negative chemical ionization options.

Key specifications include a mass range of 10-1000 Da with unit mass resolution and scan rates up to 10,000 Da/s. The instrument features an off-axis detector design and includes standard turbomolecular pumps with optional oil-free foreline configurations.

What's inside

  • Overview
  • Mass Analysis
  • Scanning Modes
  • Ionization
  • Operating Temperatures
  • Detection System
  • Vacuum System
  • Software
  • Performance Specifications
  • Physical Specifications

Specifications

Mass range10-1000 Da in 0.1 Da steps
ResolutionUnit mass resolution over the entire mass range
Scan rate5000 or 10000 Da/s
Mass axis stability+0.1 Da over 72 hr
Pumping speed (Standard Foreline Pump)greater than 95 L/min
Power Requirements (MS)90-130 Vac, 60 Hz + 3 Hz, 12 A
Weight (MS)42 Kg / 93 lbs

Frequently asked questions

What is the mass range of the Agilent 240 MS?

The mass range is 10-1000 Da in 0.1 Da steps.

Does the Agilent 240 MS support MS/MS?

Yes, MS/MS is available in all modes of operation.