Agilent Hi-Plex Ligand Exchange Columns Data Sheet

This document provides information about the usage and maintenance of Agilent Hi-Plex Ligand Exchange columns. High purity and degassed mobile phases are required. Mobile phases with up to 5% alcohol or 30% acetonitrile can be used. Sample preparation involves filtration and removal of protein, lipid, and other biological contaminants. Regular column efficiency testing is recommended.


Agilent Hi-Plex Ligand Exchange Columns Data Sheet - cover page
Brand
Agilent
Document type
Datasheet
Language
English
Pages
2
File format
PDF
File size
5.9 MB
Published
04 July, 2012
Updated
28 September, 2023
MD5 checksum
2BF26F3F68E479E422838A16BF5CDD88

About this manual

This document provides technical specifications and operational guidelines for Agilent Hi-Plex ligand exchange chromatography columns. It covers installation instructions, including the use of industry-standard end fittings and PEEK fittings to prevent dead volume. The guide details column conditioning protocols, emphasizing the need to run at elevated temperatures with a low flow rate before increasing to operating levels.

What's inside

  • Installation Column Connections
  • Shipping Eluent
  • Column Conditioning
  • Mobile Phases
  • Sample Preparation
  • Maintenance Column Efficiency Testing
  • Column Clean-up
  • System Shutdown

Specifications

Max Pressure50 bar
Warranty Period60 days

Frequently asked questions

What should be done if the column end fittings loosen?

Nuts should be tightened finger tight and then 1/4 turn using a wrench only on the outer column end fitting.

Can Hi-Plex columns be stored in a refrigerator?

Yes, they may be stored at 4 °C but under no circumstances should they be allowed to freeze.

What is the maximum alcohol content allowed in the mobile phase?

Up to 5% v/v alcohol can be used in the mobile phase.