Agilent 1100 Series Purification System Application Note: Combinatorial Library Purification

This Application Note describes the peak-based purification of 15 samples, which model a part of a combinatorial chemistry library, using the Agilent 1100 Series purification system. The samples were purified using a generic method with a gradient from 5 to 95 % organic phase in relatively short time. The re-analysis of the fractions was done on an analytical HPLC system to check the collected fractions for purity of the desired compounds.


Agilent 1100 Series Purification System Application Note: Combinatorial Library Purification - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
67 KB
Published
03 July, 2012
Updated
01 October, 2023
MD5 checksum
A75EF199045E50B8336A24947FFA0880

About this manual

Combinatorial chemistry is frequently used in drug discovery to synthesize large numbers of compounds. This document describes a method for purifying these samples using the Agilent 1100 Series purification system, leveraging the high purity typical of solid-phase synthesis libraries.

The application details a peak-based fraction collection strategy using an UV-visible detector (DAD) or mass selective detector (MSD). A generic HPLC method with a gradient from 5% to 95% organic phase was employed to separate major compounds from byproducts. The collected fractions were subsequently re-analyzed on an analytical Agilent 1100 Series HPLC system to verify purity.

What's inside

  • Introduction
  • Equipment
  • Peak-based fraction collection
  • Results and Discussion
  • Fraction re-analysis
  • Conclusion

Frequently asked questions

What is the purpose of this application note?

The document describes the peak-based purification of 15 samples that model a part of a combinatorial chemistry library using the Agilent 1100 Series purification system.

What are the peak detection criteria for the UV-visible detector?

The peak detection criteria are a threshold of 150 mAU, a peak slope of 30 mAU/min, and a maximum fraction duration of 0.5 min.