Agilent SPE Method for Anabolic Steroids Application Note

This document describes the solid phase extraction method for anabolic steroids using the AccuBOND II ODS (C18) 6 mL/500 mg cartridge. The sample is processed by adding methanol and water, followed by adsorption, elution, and concentration using vacuum. The residue is dissolved and injected into a gas chromatograph for analysis.


Agilent SPE Method for Anabolic Steroids Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
417 KB
Published
03 July, 2012
Updated
29 September, 2023
MD5 checksum
81F475FF65ED8F20E70B8F50ED156666

About this manual

This document provides a Solid Phase Extraction (SPE) method for analyzing anabolic steroids in urine and serum samples. The procedure outlines the conditioning, loading, rinsing, and elution steps using AccuBOND ODS (C18) cartridges, along with specific equipment and reagent requirements.

The document also includes a list of target steroid compounds with their partition ratios and recovery percentages, as well as the specific Gas Chromatography (GC) conditions used for the analysis, including column specifications, temperature programming, and detector settings.

What's inside

  • Condition
  • Load
  • Rinse
  • Elution
  • Equipment
  • Reagents
  • Target Compounds
  • GC Conditions
  • Column Specifications
  • Detector Settings

Specifications

Cartridge TypeAccuBOND ODS (C18) 6 mL/ 500 mg
Column TypeDB-1
Column Dimensions30 m x 0.25 mm I.D.
Film Thickness0.1 pm
Carrier GasHelium at 40 cm/ sec
Injector Temperature275°C
Detector TypeFID
Detector Temperature300°C

Frequently asked questions

What reagents are required for the SPE method?

The method requires HPLC grade water and HPLC grade methanol.

What is the recommended centrifugation speed for the cartridge?

The cartridge should be centrifuged at 1000-1500 rpm for 5 minutes.

What is the maximum temperature for concentrating the eluant?

The residue should be concentrated to dryness using gentle heat less than 45°C.