Agilent 1100 Series Capillary HPLC with Diode-Array Detection Application Note

High Sensitivity Capillary HPLC with Diode-array Detection


Agilent 1100 Series Capillary HPLC with Diode-Array Detection Application Note - cover page
Brand
High
Category
Other
Document type
Product Leaflet
Language
English
Pages
2
File format
PDF
File size
62 KB
Published
03 July, 2012
Updated
30 September, 2023
MD5 checksum
D49946BA8584D0D9CB51DEE4CF38344F

About this manual

This document describes how to perform high-sensitivity capillary HPLC analysis using the Agilent 1100 Series HPLC high pressure gradient system. It details the prerequisites for running capillary columns, including the need for low system delay volume and specific detector cells.

What's inside

  • Abstract
  • Prerequisites
  • Table 1 Comparison
  • Conversion of Standard System
  • Peptide Identification
  • Figure 2 Peptide Map
  • Conditions
  • Authors
  • Copyright

Specifications

Column internal diameter180 to 300 µm
Flow rates2 to 4 µL/min
Column length100 to 200 mm
Flow cell volume500 nL
Flow cell pathlength10 mm
Analysis rangefemtomole range

Frequently asked questions

What prerequisites are needed to run capillary columns on standard HPLC equipment?

The system requires low system delay volume, the addition of a flow splitter, and the availability of nL detector cells.

What is the advantage of diode-array detection for peptide identification?

It allows spectra of eluting compounds to be stored digitally and manipulated by algorithms, which is useful for identifying peptides with aromatic amino acids.