Screening for 171 Volatile Organic Air Pollutants Using GC/MS with Deconvolution Reporting Software and a New Indoor Air

This document describes an indoor air analysis method using thermal desorption/gas chromatography/mass spectroscopy (TD/GC/MS), and the creation of an Indoor Air Toxics Database. The database contains retention times and mass spectra for 171 volatile and semi-volatile organic compounds. With the use of the database and Agilent's Deconvolution Reporting Software (DRS), it is possible to identify these compounds within 2 to 3 minutes after TD/GC/MS analysis. The document also discusses the features and capabilities of DRS, as well as the analysis results of air samples from an office building and a carpet warehouse.


Screening for 171 Volatile Organic Air Pollutants Using GC/MS with Deconvolution Reporting Software and a New Indoor Air - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
16
File format
PDF
File size
614 KB
Published
02 July, 2012
Updated
28 September, 2023
MD5 checksum
59EFFBC05A27775840E9E258F2CFF3AB

About this manual

The document presents a method developed by Agilent Technologies and Markes International for analyzing indoor air pollutants. It utilizes a retention-time-locked mass spectral database containing 171 volatile and semivolatile organic compounds, specifically designed for use with the Agilent Deconvolution Reporting Software (DRS). The method allows for the identification of these compounds in just 2 to 3 minutes following thermal desorption/gas chromatography/mass spectroscopy analysis.

The technique employs a new Indoor Air Toxics Database combined with DRS capabilities, such as deconvoluting overlapping mass spectra to identify unresolved GC peaks reliably. The application details the instrumentation used, including the Markes UNITY Thermal Desorber and Agilent 6890N Gas Chromatograph coupled with an Agilent 5975 Mass Selective Detector, along with specific software configurations like ChemStation, AMDIS, and NIST MS Search.

What's inside

  • Introduction
  • Instrumentation and Conditions of Analysis
  • Sample Collection
  • Results and Discussion
  • Deconvoluted peaks and spectra

Frequently asked questions

How long does it take to identify compounds after data acquisition?

It is possible to identify any of the target compounds in just 2 to 3 minutes after a TD/GC/MS analysis.

What size air samples were collected for the office building test?

Six- or twelve-liter air samples were collected by pumping at 200 mL/min.

How many compounds were identified in the office air sample?

DRS was able to identify more than 80 compounds in the office air.