Agilent Technologies 7500c ICP-MS RoHS/ELV Heavy Metals Measurement Application Note

This application note describes a robust analytical method for the determination of heavy metals and sub- substances regulated by the EU Restriction of Hazardous Substances/End of Life Vehicles (RoHS/ELV) directives. Based on the experience of Nihon Environmental Services Co., Ltd in Japan using an Agilent Technologies 7500c ORS (Octopole Reaction System) collision/reaction cell ICP-MS, this note summarizes: • An overview and validation of ICP-MS methodology for RoHS applications • Benefits of ICP-MS methodology to the analysis of trace elements in high matrix samples and limitations of EDXRF (Energy Dispersive X-Ray Fluorescence) for the measurement of lead in flame resistant plastics • Advantages of ICP-MS for RoHS applications.


Agilent Technologies 7500c ICP-MS RoHS/ELV Heavy Metals Measurement Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
179 KB
Published
02 July, 2012
Updated
30 September, 2023
Model
7500c
MD5 checksum
D262D1067DF5A91D7275305D9499E2BC

About this manual

This document outlines the use of Inductively Coupled Plasma Mass Spectrometry (ICP-MS) to measure heavy metals such as cadmium, lead, mercury, and chromium in compliance with EU RoHS and ELV directives. It details a validated methodology using an Agilent Technologies 7500c ICP-MS equipped with an Octopole Reaction System (ORS) collision/reaction cell.

What's inside

  • RoHS/ELV Directives
  • ICP-MS Methodology
  • Operating Parameters
  • Isotopes and Internal Standards
  • Method Validation
  • High-Concentration Matrix Samples
  • Managing Interferences
  • EDXRF Limitations

Frequently asked questions

How does the Agilent 7500c handle spectral interferences?

Spectral interferences are overcome by operating the ORS collision/reaction cell in helium mode with a flow rate of 4.5 mL/min.

What flow rate is recommended for the helium mode in the ORS collision/reaction cell?

The helium mode should be operated at a flow rate of 4.5 mL/min.

What is the certified cadmium value for the standard polyethylene substance CBR-680?

The certified value for cadmium in this standard is 140.8 mg/kg.