Agilent SampliQ OPT Tetracycline Determination Application Note

A simple and effective method has been developed for the simultaneous determination of the seven tetracycline antibiotic residues in chicken meat.


Agilent SampliQ OPT Tetracycline Determination Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
345 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
9DBC43C669E2D9C67631341E6625B23E

About this manual

This document details a validated method for the simultaneous determination of seven antibiotic residues (minocycline, oxytetracycline, tetracycline, demeclocycline, chlortetracycline, methacycline, and doxycycline) in chicken tissue. The method utilizes solid-phase extraction (SPE) with Agilent SampliQ OPT cartridges followed by high-performance liquid chromatography (HPLC) with UV detection.

The procedure covers sample preparation using EDTA-Mcllvaine buffer, SPE purification steps, and specific HPLC conditions including column type, flow rate, and gradient elution. Validation results demonstrate acceptable recoveries ranging from 59.0% to 99.0% and limits of detection between 2.5 and 5 µg/kg, meeting Chinese regulatory standards.

What's inside

  • Introduction
  • Materials and Chemicals
  • HPLC Conditions
  • Sample Preparation
  • SPE Purification
  • Results and Discussion
  • Recovery and Reproducibility
  • Conclusions

Specifications

Column dimensions250 mm x 4.6 mm id
Flow rate1.5 mL/min
Detector wavelength350 nm
Injection volume100 µL
Limits of detection2.5 to 5 µg/kg
Linear calibration range25 to 500 µg/kg
Recovery range59.0% to 99.0%

Frequently asked questions

Which tetracyclines are determined in this method?

The method determines minocycline, oxytetracycline, tetracycline, demeclocycline, chlortetracycline, methacycline, and doxycycline.

What SPE cartridge is used for purification?

The method uses an Agilent SampliQ OPT cartridge.

What is the column temperature for the analysis?

The column temperature is set to 30 °C.

What is the mobile phase composition?

The mobile phase consists of methanol, acetonitrile, and 10 mmol/L TFA solution with gradient elution.