Agilent 6460 Triple Quadrupole LC/MS Amino Acid and Acylcarnitine Analysis Application Note

This application note describes a rapid method for the simultaneous detection and quantification of 37 acylcarnitines and 12 amino acids in dried blood spots using the Agilent Triple Quadrupole LC/MS System in neutral loss scan, precursor ion scan and multiple reaction monitoring (MRM) data acquisition modes. With a run time of 1.6 minutes, the assay provides the ability to analyze 400-500 samples per day, and custom reports provide all the information needed to rapidly list the critical elements for the analysis.


Agilent 6460 Triple Quadrupole LC/MS Amino Acid and Acylcarnitine Analysis Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
12
File format
PDF
File size
167 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
FF863264CCC8179E80C2A2750FE5E07D

About this manual

This document details a clinical research method for the simultaneous analysis of 37 acylcarnitines and 12 amino acids in dried blood spots (DBS). The method utilizes the Agilent Triple Quadrupole LC/MS System with a run time of 1.6 minutes, allowing for the processing of 400-500 samples per day. The assay employs neutral loss scan, precursor ion scan, and multiple reaction monitoring (MRM) data acquisition modes to detect and quantify the analytes.

The procedure covers sample preparation involving extraction with methanol containing isotopic internal standards, filtration, derivatization with butanol-HCl, and reconstitution. The document provides specific instrument conditions for the Agilent 1200 Series LC system coupled with an Agilent 6460 Triple Quadrupole LC/MS, including mobile phases, flow rates, and mass spectrometer settings for positive ESI mode.

What's inside

  • Authors
  • Introduction
  • Reagents and standards
  • Sample preparation
  • LC Run Conditions
  • Neutral Loss Scan
  • Precursor Ion Scan
  • Multiple Reaction Monitoring
  • Data Analysis

Specifications

Run time1.6 minutes
Samples per day400-500
Acylcarnitines analyzed37
Amino acids analyzed12
Injection volume5 ul
Flow rate0.6 mL/min
Drying gas temperature300 °C
Capillary voltage4000 V

Frequently asked questions

What is the run time for the assay?

The run time for the assay is 1.6 minutes.

How many samples can be analyzed per day?

The assay provides the ability to analyze 400-500 samples per day.

Which data acquisition modes are used?

The method uses neutral loss scan, precursor ion scan, and multiple reaction monitoring (MRM) modes.