Agilent SpectrAA-40P Arsenic Determination Application Note

This method uses a vapor generation accessory to determine arsenic in non-silicate geological ore samples.


Agilent SpectrAA-40P Arsenic Determination Application Note - cover page
Brand
The
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
60 KB
Published
12 June, 2012
Updated
01 October, 2023
MD5 checksum
A1B16BE4EB8A477952B324CDFCF95E12

About this manual

The document compares two digestion methods for analyzing arsenic in geological samples: aqua regia (nitric acid/hydrochloric acid) and hydrochloric acid/hydrogen peroxide. These methods are effective only for arsenic not bound within a silica matrix.

Measurements were performed using an Agilent SpectrAA-40P spectrometer equipped with a Vapor Generation Accessory (VGA-76). The note details sample preparation, stabilization techniques using potassium iodide, and provides data on analytical ranges, calibration graphs, and recovery studies for both digestion methods.

What's inside

  • Introduction
  • Practical
  • Digestion Solutions
  • Analytical Range of Arsenic
  • Stabilization
  • Recoveries
  • Calculations
  • Results
  • Conclusion

Specifications

Spectrometer ModelAgilent SpectrAA-40P
Accessory ModelVapor Generation Accessory (VGA-76)
Sample Changer ModelProgrammable Sample Changer (PSC-56)
Analytical Range 197.2 nm0-100 ng/mL

Frequently asked questions

Can these methods extract arsenic from silica matrices?

No, neither digestion method can extract arsenic from within a silica matrix; hydrofluoric acid is required for that.

What stabilizer is recommended for final solutions?

Solid potassium iodide should be added to stabilize solutions, reducing As(V) to As(III).