Respironics Joey Dosimeter User Instructions

The Joey Dosimeter is a wide-band spectroradiometer used in neonatal phototherapy. It measures the irradiance of the wavelengths of blue light, 425nm – 475nm.


Respironics Joey Dosimeter User Instructions - cover page
Brand
The
Document type
User Manual
Language
English
Pages
2
File format
PDF
File size
494 KB
Published
03 August, 2012
Updated
30 September, 2023
MD5 checksum
88719F4839424D2AAFB47895B30DB491

About this manual

The Respironics Joey Dosimeter is a wide-band spectroradiometer designed specifically for use with neonatal phototherapy devices. It measures irradiance of blue light wavelengths between 425nm and 475nm, peaking at 450nm, in units of microwatts per centimeter squared per nanometer (uw/cm²/nm). The device features a digital display that reads values from 00.0 to 99.0 uw/cm²/nm.

What's inside

  • Description
  • Use with Fiberoptic Phototherapy Devices
  • Use with Overhead Phototherapy Devices
  • Battery Information
  • Replacing the Battery
  • Cleaning
  • Recalibration
  • Warranty

Specifications

Wavelength Range425nm — 475nm
Peak Wavelength450nm
Measurement Unituw/cm²/nm
Display Range00.0 to 99.0 uw/cm²/nm
Battery TypeSingle nine volt (9V) nonrechargeable battery
Calibration Accuracy+/-2% or +/- one digit

Frequently asked questions

What wavelengths can the Joey Dosimeter measure?

The device is designed only for measuring blue light wavelengths between 425nm and 475nm.

How do I replace the battery?

Remove the cover on the lower back side, lift out the 9-volt battery, and replace it with a new one before replacing the cover.

When should the device be recalibrated?

It is recommended that the instrument be returned for recalibration annually, or immediately if physically abused.