The Measurement of Lead in Food Products by Graphite Furnace AA SpectraAA-40 Application Note

This application note describes the use of the SpectraAA-40 and the graphite furnace atomic absorption spectrometry method for the determination of lead in foods.


The Measurement of Lead in Food Products by Graphite Furnace AA SpectraAA-40 Application Note - cover page
Brand
The
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
546 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
FDB2DA8E6EE2DA0D7EB2EB842200F92D

About this manual

The document details an analytical study measuring trace levels of lead in food products, specifically infant milk powder formula and digested fish tissue. It utilizes Graphite Furnace Atomic Absorption (GFAA) spectroscopy to detect these elements at very low concentrations.

The text compares two sample introduction methods: wall atomization using partitioned pyrolytic coated graphite tubes and platform atomization using a pyrolytic platform. Results indicate that while infant milk powder can be measured directly in solution with minimal background interference, fish digests benefit significantly from the use of a pyrolytic platform to reduce excessive background absorption.

What's inside

  • Introduction
  • Experimental Setup
  • Sample Treatment and Reagents
  • Infant Milk Powder Analysis
  • Fish Tissue Analysis
  • Instrument Parameters

Frequently asked questions

What instrument was used for the lead measurement study?

A SpectrAA 40 was used together with the Agilent GTA-96 graphite furnace atomizer and automatic sample dispenser.

What wavelength was used for the lead lamp?

A hollow cathode lead lamp was used at a wavelength of 283.3 nm.

What chemical modifier was used in the analysis?

Ammonium dihydrogen phosphate crystals were dissolved to give a 0.5% w/v solution and used as a chemical modifier.